Click on the
Scaling
tab to scale your transmitter. You will be given the choice of three scaling
methods: 1) Scale and Offset method, 2) Coordinates of 2 points method where (Low In, Low
Read) and (High In, High Read) data points are entered numerically, and 3) Reading Coor-
dinates of 2 points method, which captures actual readings.
Click on the
Filter
tab to set to set up filtering for your readings. The filter time constant can be
automatic, be specified in seconds, or be turned off. The adaptive threshold modifies the time
constant in response to noise. A low adaptive threshold is recommended for normal low noise.
A high adaptive threshold is recommended for high noise environments.